AD-4408C Weighing Indicator Page 34
8.3. Standard Serial Output (Cl f)
Function No.
Parameter or
Setting range
Function Description
Default
value
Cl f01
1
to 5
Output
data
1: Displayed weight
2: Gross weight
3: Net weight
4: Tare weight
5: Gross/Net/Tare
1
Cl f02
1
to 3
Data
transfer
mode
1: Stream
2: Auto printing
(See the NOTE below)
3: Manual printing
1
Cl f03
1 to 2
Baud rate
1: 600 bps
2: 2400 bps
2
NOTE: When “
1: Enables restriction” is selected for Calf21 (Communication
restriction), auto printing will not be performed even if auto printing
conditions are met.
8.4. CC-Link (CC f)
Function No.
Parameter or
Setting range
Function Description
Default
value
CC f01
1 to 64
Station
number
n: Station number
1
CC f02
0 to 2
Occupied
stations
0: One
1: Two
2: Four
2
CC f03
0 to 4
Baud rate
0: 156 kbps
1: 625 kbps
2: 2.5 Mbps
3: 5 Mbps
4: 10 Mbps
4
CC f04
0 to 1
Initialization
0: Not required
(When the AD-4408C is turned on
and the CC-Link is ready, the remote
READY flag (RX007B) becomes
active without performing initialization.)
1: Required
(When the AD-4408C is turned on
and the CC-Link is ready, the remote
READY flag (RX007B) becomes
active after the initialization is
complete.)
1
CC f05
0 to 2
Output
data
0: Displayed weight
1: Net weight
2: Gross weight
Specify what data to output when the
number of occupied stations is 1 or 2.
0