Page 23 AD-4408C Weighing Indicator
7.3. Calibration-Related Functions (Calf)
Step 1 Press the ON/OFF key to turn the indicator ON (weighing mode) if it is in the
OFF mode.
Remove the CAL key cover and press the CAL key.
Cal is displayed to indicate that the indicator enters the calibration mode.
Step 2 Press the ENTER key to display Cal5et . The indicator enters the mode of
calibration with an actual load. To return to the weighing mode, press the ESC
key.
Step 3 Press the ∧ or ∨ key to select CalfF and press the ENTER key.
Step 4 Press the ∧ or ∨ key to select a function number to be used and press the
ENTER key. The current setting value is displayed.
Step 5 Change the value as necessary. Two methods to change are available,
depending on the function number as below
Method Description
Selecting a parameter
Only the available parameter is displayed and blinks.
Press the ∧ or ∨ key to select a parameter.
Inputting the value
All the digits are displayed and a digit to be changed
blinks.
Press the < or > key to select a digit and
press the ∧ or ∨ key to change the value.
After setting, press the ENTER key. The next function number is displayed.
When the value is not to be changed, press the ESC key to return to the
function number display.
Step 6 Press the ESC key. Cal5et is displayed and the data of the
calibration-related functions is written into the EEPROM.
The current state is the same as that of step 2. To return to the weighing mode,
press the ESC key again.
NOTE: If a value exceeding the settable range is entered, err dt is displayed and
the input is canceled.
If the decimal point blinks, it indicates that the current value is not the
weight value.