13 ANALOG OUTPUT (AD-8922A-06)
Converts the specified digits of the weighing data received from the weighing instrument to voltage and
outputs. The output voltage can be selected from "0 to 1 V" and "0.2 to 1 V".
Note: When the AD-8922A-06 is installed, the pin assignment (DIN 8-pin), of the RS-232C serial
interface, to connect an external device, will be changed.
13-1 Analog Output Specifications
Accessories
DIN connector (Plug) 1 pc.
Screwdriver 1 pc.
Instruction manual 1 copy
Panel view
Analog output
Output impedance 100 Ω or less
Linearity 0.3% or less
Output range 0 V-1 V (With the slide switch set to “0V~“)
0.2 V-1 V (With the slide switch set to “0.2V~“)
RS-232C
Transmission system : EIA RS-232C
Transmission form : Asynchronous, bi-directional, half duplex
Data format : Baud rate : 600, 1200, 2400, 4800, 9600, 19200 bps
Data bits : 7 bits or 8 bits
Parity bit : EVEN, ODD (Data bits 7 bits)
NONE (Data bits 8 bits)
Stop bits : 1 bit or 2 bits
Code : ASCII
Terminator : <CR> or <CR><LF>