A&D AD-8922A Bedienerhandbuch

Stöbern Sie online oder laden Sie Bedienerhandbuch nach Fernbedienungen A&D AD-8922A herunter. A&D AD-8922A Instruction manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - Production Weighing Unit

AD-4212C-300 AD-4212C-600 AD-4212C-3000 AD-4212C-6000 AD-4212C-301 AD-4212C-3100 Production Weighing Unit INSTRUCTION MANUAL 1WMPD4001992B

Seite 2

8Rear of theweighing unitRS-232C cableD-sub 9-pinArrow markface up2.2. Installing the Balance Install the balance as follows: 1. Refer to “3. PRECAU

Seite 3 - CONTENTS

9MASS OF THE LOCALLY FABRICATED WEIGHING PAN Design the weighing pan so that the mass falls within the ranges shown in the table below: Model Statu

Seite 4 - 1. INTRODUCTION

103. PRECAUTIONS To get the optimum performance from the balance and acquire accurate weighing data, note the following: 3.1. Before Use  Install

Seite 5

113.2. When Building Into a System The AD-4212C is a precision balance. When it is built into a system, errors such as unstable weight values may o

Seite 6

12 Errors due to a static charge and measures to take When the ambient humidity is less than 45% RH, insulators such as plastic or glass are prone

Seite 7 - 1.2. Compliance

13 Errors due to air flow and measures to take Where the influence of ambient air flow is great such as: close to an air conditioner, door or pass

Seite 8 - 2. UNPACKING THE BALANCE

143.3. During Use  To minimize the affect by electrical noises, earth ground the weighing unit and the display unit (option).  Do not drop things

Seite 9 - Leveling foot

154. OPERATION OF WinCT-AD4212C The software has the following functions: - Set the weighing speed - Set the minimum display value - Calibration

Seite 10 - 2.2. Installing the Balance

164.1. Changing the Weighing Speed Using WinCT-AD4212C The weighing speed can be selected from the following three rates to minimize the influence

Seite 11

17 Calibration procedure This function calibrates the balance using the calibration weight. (Display example: AD-4212C-300) Operation 1. Plug in t

Seite 12 - 3. PRECAUTIONS

© 2011 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into

Seite 13

184.3. Changing the Baud Rate Using WinCT-AD4212C Operation 1. Click the [Settings] tab. 2. Select the baud rate to be used and click the Set

Seite 14 - Grounding

195. SMART RANGE FUNCTION The AD-4212C-301/3100 are equipped with two ranges. The precision range has a higher resolution. The standard range has a n

Seite 15 - Magnetic material

206. COMMUNICATION SPECIFICATIONS The AD-4212C series can communicate interactively using RS-232C. The AD-4212C series can continuously transmit th

Seite 16 - 3.5. Power Supply

212. Pin position Pin position of the RS-232C cable (D-Sub 9-pin side, female) Pin No. Signal name Direction Description 1 (Vs) - Internally u

Seite 17 - Before use

224. Command The balance can be controlled by the following commands. When transmitting, add the terminator <CR><LF> to each command. Co

Seite 18 - 4.2. Calibration

23 CAL command example (PC, PLC)External devices Time Weighing unit  CAL commandDisplays

Seite 19 - Operation

24 SMP command example R command example when the precision range is used SMP commandSMPCRLFST,+0000.012 gS

Seite 20 - Set button

257. MAINTENANCE  Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.  Use the original pac

Seite 21 - 5. SMART RANGE FUNCTION

268. TROUBLESHOOTING 8.1. Checking the Balance Performance and Environment The balance is a precision instrument. When the operating environment or

Seite 22 - 1. RS-232C specification

27 Sample and container  Has the sample absorbed or lost moisture?  Has the temperature of the sample, and the container if used, been allowed to

Seite 23 - 3. Data output format

1CONTENTS 1. INTRODUCTION...2

Seite 24 - R command example

289. SPECIFICATIONS AD-4212C-300 AD-4212C-600 AD-4212C-301 Weighing capacity 320 g 620 g 320 g / 51 g*1 Maximum display 320.084 g 620.084 g

Seite 25 - CAL command example

29 AD-4212C-3000 AD-4212C-6000AD-4212C-3100 Weighing capacity 3200 g 6200 g 3200 g / 510 g*1 Maximum display 3200.84 g 6200.84 g 3200.84 g / 51

Seite 26 - SMP command example

3010. EXTERNAL DIMENSIONS - AD-4212C-300 / 600 / 3000 / 6000 / 301 / 3100 Unit: mm 9848.83018.846505091.9 1

Seite 27 - 7. MAINTENANCE

3111. OPTIONS AX-KO3590-200 RS-232C cable 2 m AX-KO3590-500 RS-232C cable 5 m AX-KO3590-1000 RS-232C cable 10 m (provided with the balanc

Seite 28 - 8. TROUBLESHOOTING

32 AD-1684: Electrostatic field meter  This option measures the amount of the static charge on the sample, tare or peripheral equipment and display

Seite 29 - 8.2. Asking For Repair

3312. TERMS/INDEX Terms Stable value The weight value when the stabilization indicator appears. Environment Ambient conditions such as vibration, d

Seite 30 - 9. SPECIFICATIONS

34AX- KO3590-500 ...4, 31 AX- KO3590-1000 ...4, 31 AX-USB-9P...

Seite 31

35Remote controller ...31 Repeatability... 26, 28, 29, 33 RE-ZERO button ...

Seite 32 - Pan support

21. INTRODUCTION This manual describes how the AD-4212C series balance works and how to get the most out of it in terms of performance. Read this

Seite 33 - 11. OPTIONS

3 Example 1 (Connecting to a personal computer or an AD-8922A) Accessory RS-232C cable (10 m)PersonalcomputerU

Seite 34

4 Example 2 (Connecting to a PLC) . * Specify the AD-8922A-01 connection cable when ordering the AD-8922A-01. AX-KO3705-200 2 m

Seite 35 - 12. TERMS/INDEX

51.2. Compliance Compliance with FCC Rules Please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has

Seite 36

6Windows communicationtools (WinCT-AD4212C)AC adapterID labelRS-232C cable (10 m)Weighing panPan supportScrews (Flat head: M4 X 6)Calibration weight

Seite 37

7ATTACHING THE AD-4212C SERIES WEIGHING UNIT DIRECTLY ON THE MOUNTING BASE When the AD-4212C series weighing unit is built into a system, remove the

Kommentare zu diesen Handbüchern

Keine Kommentare