A&D AD-8922A Bedienerhandbuch Seite 22

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 37
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 21
20
6. COMMUNICATION SPECIFICATIONS
The AD-4212C series can communicate interactively using RS-232C.
The AD-4212C series can continuously transmit the weighing data. (Approx.10 times/second* to 50
times/second)
The AD-4212C series is DCE. The AD-4212C series can connect directly to the RS-232C terminal of a
personal computer by using the accessory RS-232C cable.
* Factory setting
1. RS-232C specification
Transmission system EIA RS-232C
Transmission form Asynchronous, bi-directional, half duplex
Transmission rate Approx.10 times/second* to 50 times/second
Data format Baud rate 2400bps* to 19200bps
Data bits 7 bits
Parity EVEN
Stop bit 1 bits
Code ASCII
Terminator <CR><LF>
* Factory setting
Relation between baud rate and transmission rate of weighing data
Baud rate
Transmission rate of weighing data
2400bps 12.5 times/second
4800bps 25.0 times/second
9600bps 50.0 times/second
19200bps 50.0 times/second
1
0
-5V
to -15V
+5V
to +15V
Data bits Start bit
Parity bit
Stop bit
LSB MSB
Seitenansicht 21