A&D GR Series Instruction Manual
GB 50
When weighing data is being placed in memory, the data cannot be output to the RS-232C
interface.
The “FUL” displays full memory. More data cannot be stored until deletion of the stored data.
Automatic self calibration cannot be used while the interval memory mode is working.
The following commands cannot be used during data storage.
Q Query command for weighing data.
S Request command for stable weighing data.
SI Query command for weighing data.
SIR Request command for continuous weighing data.
11.2. Preparation of the Function Table
Item Data output Auto print Data memory Interval time
Mode mode mode mode
Key mode Prt 0 —– dAtA 1
Auto-print mode A Prt 1 AP-P 0 ~ 2 dAtA 1 —–
Auto-print mode B Prt 2 AP-b 0 ~ 2 dAtA 1
Interval Memory Mode Prt 3 —– dAtA 1 int 0 ~ 8
Not used data number d-no 0
Use data number d-no 1
Note
The data memory function does not work
with dAtA 0.
11.3. Output of Data from Memory
Displaying and Transmitting the Data
1. Press and hold the [PRINT] key until displaying rECALL and release the key.
2. Press the [PRINT] key to enter the mode. Use the following keys.
[RE-ZERO] key The key to proceed to the next data.
[MODE] key The key to go back to the previous data.
[PRINT] key The key to transmit the current data to the RS-232C interface.
[RANGE] key is pressed and The key to delete the current data.
held then press the [CAL] key
[CAL] key The key to exit this mode.
3. Press the [CAL] key. The balance returns to weighing mode.