8.2. Details of the Function Table
English
GB 35
Class Item Parameter Summaries
bASFnc Cond 0 Fast response, Sensitive value Common data of “Response
Environment, Condition • 1 adjustment”.
Display 2 Slow Response and stable value
St-b 0 Stable when within ±1 digit The stability indicator lights when
Stability band width • 1 the display fluctuation is within
2 Stable when within ±3 digit the range per second
trc 0 OFF The function to keep zero display
Zero tracking • 1 ON by tracking zero-drift.
SPd • 0 Normal, 5 times/second The period to refresh the display.
Display update rate 1 Fast, 10 times/second
Pnt • 0 Point (.) The form of decimal point.
Decimal point 1 Comma (,)
P-on • 0 OFF Connecting adaptor, the display
Automatic start 1 ON turns on wuthout key operation.
dout • 0 Key mode Data is output or stored with
Data output PRINT key and stability indicator.
1 Auto-print mode A Data is output or stored when the
Prt (Standard value is zero) display value is stable and meets
Data output mode Auto print mode B the conditions of AP-P, AP-b
2 (Standard value is last and standard value.
stable value)
In case of dAtA 0,
3 Stream mode/ Data is output continuously.
Interval memory mode In case of dAtA 1,
Data memory function is used.
AP-P • 0 Plus polarity Display value * Standard value
Auto-print polarity 1 Minus polarity Standard value > Display value
for mode A or B 2 Both polarities (Absolute value) Display value * Standard value or
Standard value > Display value
AP-b 0 10 digit Difference between standard
Auto-print difference • 1 100 digit value and display value.
for mode A or B 2 1000 digit
dAtA • 0 Not used Relation: Prt, int, d-no
Data memory function 1 Use
• 0 Every Measurment
1 Every 2 seconds Interval time is selected on
int 2 Every 5 seconds Prt 3
, dAtA 1.
Interval time for 3 Every 10 seconds
Data memory function 4 Every 30 seconds
5 Every 1 minute
6 Every 2 minutes
7 Every 5 minutes
8 Every 10 minutes
•: factory setting.
*: “Digit” is the unit of minimum display.