5.3. Operation by Commands
5.3.1. Handling Commands
Specify the R / W flag either to the read command or the write command.
R / W flag: 0: Write command 1: Read command
Specify a command to execute to the command No.
Specify the command data to be written to Write data.
The command will be effective at the rising edge of the command request flag.
The signal level must be maintained for 30 msec minimum.
The response result to the command request is output as the command request
response flag.
The command response result is output as the command No. response.
The response result of the read command is output as the read data.
5.3.2. Command execution procedure
Before execution
Step 1 Confirm that the command request flag is turned OFF.
Step 2 Specify the R / W flag either to the read command or the write command.
R / W flag: 0: Write command 1: Read command
Step 3 Specify the command to execute to the command No.
Step 4 Specify the command data to be written to Write data when necessary.
Execution
Step 5 Confirm that the slave ready bit is turned ON.
Step 6 Turn the command request flag ON. The command will be executed at the rising
edge.
Step 7 The AD-4408A responses. The response results are output as the command
request response flag, R / W response flag and command No. response.
Step 8 When the read command is specified by the R / W flag, a response will be output
as the read data.
After execution
Step 9 Turn the command request flag OFF.