6. Operation
6-1 Turning on the POWER switch
Turn the POWER switch on. After several seconds of self-check, the home display appears.
Note
When the self-check detects any error, the error code is displayed. The error code display of some
errors is frozen. Refer to "7-6 Error Codes" for how to solve the error.
The self-check checks the data in memory. The duration of the self-check depends on the number of
memory data.
6-2 Attaching the Cuff/ECG Electrodes/SpO
2
Sensor
Consider "Safety Precautions" appearing in an earlier part of the instruction manual, when attaching the
Cuff/ECG electrodes/SpO
2
sensor.
Attaching the cuff
1. Hold the cuff with the hose pointing
toward the lower part of the arm.
2. Wrap the cuff around the upper arm,
about 3 5 cm above the elbow, with the
ARTERY mark on the artery, as shown in
the illustration.
3. Allow some slack between the cuff and
arm so that a finger or two can be
inserted.
4. When the cuff inner end is not within the
area labeled RANGE, the cuff size is not
appropriate. Replace the cuff.
Note
Constriction of the upper arm, caused by rolling up a shirt sleeve, may prevent accurate readings.
When the size of the supplied cuff does not fit the patient, refer to "8-1 Accessories/Options" to find the
proper sized cuff.