The trigger makes repeating waveforms appear static
on the oscilloscope display by repeatedly displaying
the same portion of the input signal. Imagine the
jumble on the screen that would result if each sweep
started at a different place on the signal (see Figure
34).
Trigger Level and Slope
Your oscilloscope may have several different types of
triggers, such as edge, video, pulse, or logic. Edge
triggering is the basic and most common type.
For edge triggering, the trigger level and slope
controls provide the basic trigger point definition.
The trigger circuit acts as a comparator. You select
the slope and voltage level of one side of the
comparator. When the trigger signal matches your
settings, the oscilloscope generates a trigger.
• The slope control determines whether the trigger
point is on the rising or the falling edge of a
signal. A rising edge is a positive slope and a
falling edge is a negative slope.
• The level control determines where on the edge
the trigger point occurs.
Figure 35 shows the effect the trigger slope and level
settings have on how a waveform is displayed.
Trigger Sources
The oscilloscope does not necessarily have to trigger
on the signal being measured. Several sources can
trigger the sweep:
• Any input channel
• An external source other than the signal applied
to an input channel
• The power source signal
• A signal internally generated by the oscilloscope
Most of the time, you can leave the oscilloscope set
to trigger on the channel displayed. Many oscillo-
scopes provide a trigger output that delivers the
trigger signal to another instrument.
Note that the oscilloscope can use an alternate
trigger source whether displayed or not. So you have
to be careful not to unwittingly trigger on, for
example, channel 1 while displaying channel 2.
Trigger Modes
The trigger mode determines whether or not the
oscilloscope draws a waveform if it does not detect a
trigger. Common trigger modes include normal and
auto.
In normal mode, the oscilloscope only sweeps if the
input signal reaches the set trigger point; otherwise
(on an analog oscilloscope) the screen is blank or (on
a digitizing oscilloscope) frozen on the last acquired
waveform. Normal mode can be disorienting since
you may not see the signal at first if the level control
is not adjusted correctly.
Auto mode causes the oscilloscope to sweep, even
without a trigger. If no signal is present, a timer in
the oscilloscope triggers the sweep. This ensures that
the display will not disappear if the signal drops to
small voltages. It is also the best mode to use if you
are looking at many signals and do not want to
bother setting the trigger each time.
Figure 34. Untriggered display.
Figure 35. Positive and negative slope triggering.
19
Kommentare zu diesen Handbüchern