Setting
CALF- Name User’s
Default Parameter Description
0
1
Does not output when the weight value is overflowing
or unstable
0
12
Standard Serial
Output; Output
When Weight Value
Is Overflowing or
Unstable
1
1
Outputs even if the weight is overflowing or unstable
0
Does not output when the weight value is overflowing
or unstable.
13
RS-232C/-422/-485
; Output When
Weight Value Is
Overflowing or
Unstable
1
1
Outputs even if the weight value is overflowing or
unstable
1
Normal batching
(Customer
programmed
control mode)
2
Loss-in-weight
(Customer
programmed
control mode)
3
Normal batching
(Built-in
automatic program
mode)
4
Loss-in-weight
(Built-in
automatic program
mode)
Batch
Weighing
For the hopper
scale use
5 Check weighing 1
6 Check weighing 2
7 Check weighing 3
14 Weighing Mode 3
8 Check weighing 4
Check-Wei
ghing
For the platform
scale use
15 Zero Input Voltage 0.000000
0.000000
to
2.200000
Input Voltage (mV/V) from the Load Cell at "Zero",
which is determined in "Zero Calibration" with weights.
16
Span Input Voltage
(Capacity to zero)
3.200000
0.000000
to
3.200000
Input Voltage (mV/V) from the Load Cell at "Span",
which means the difference between Capacity and
Zero. The voltage is determined in "Span Calibration"
with weights.
17
Weight against
Span Input Voltage
16000
0 to 800000
(decimal
point
ignored)
When "Digital-Calibration", calibration not using
weights, is performed, CALF-15 and CALF-16 also
have to be done.
This setting will be changed to Capacity (CAL-04) if
"CAL Set" is done.