Page 26 AD-4406 Weighing Indicator
Data Output
Item
Parameter
Description
* 0 No data output
1 Analog output Set F31, 32, 33
2 Serial output
F30
Data output
3 Serial output (Zero suppressing)
RS-232C,
RS-422/485
*: Initial settings (In case of installing RS-232C, F30 is set to 2. )
Analog Output
Item
Parameter
Description
* 0 Displayed data
1 Gross data
F31
Output data
2 Net data
F32
Weight value at 4 mA output
-999999 to 999999
(Initial setting is 0)
F33
Weight value at 20 mA output
-999999 to 999999
(Initial setting is 10000)
Polarity is changed
by pressing the +/-
key.
*: Initial settings
Serial Data Format
Item
Parameter
Description
x 0 Terminator
x 1 Device ID (selected at F06)
x 2 Code number
x 3 Data number
x 4 Result of comparison
x 5 (Reserved)
x 6 (Reserved)
x 7 Accumulated value
x 8 Accumulation count
x 9 Stable/Over
x A Displayed weight
x B Gross weight
x C Net weight
x D Tare weight
x E
Weight type (G /N / T, refer to
CF06)
F34
Serial data format 1
Initial value
19」,2E」,3A,4F,E50
F35
Serial data format 2
Initial value
17,2F,E30
x F Weight unit (refer to CF07)
100,10:order of
output (1-999)
Select with the 1 1
key.
1: output data
Select with the
key.
"Exxx" indicates
the end of data to
output.
Press
NET/GROSS to
expand or shorten
the output data length.
1 “」” (right hand of the figure) of the display indicates that the comma (F46=0, semi-colon
for F46=1) is output after the data. To turn this mark on and off, press the ZERO and
TARE keys at the same time.
Output of initial settings of F34 (19」, 2E」, 3A, 4F, 50) is like “ST,GR,12345.6kg”.
2 The data number (parameter 3) increments automatically with each serial data output
from 1 to 99999 (the next of 99999 is 1). The starting number can be set by the keys
(F2 + NET/GROSS).