7. CALIBRATION
The AD-4531B measures voltage signals from sensors and displays the
values. Calibration is performed so that the AD-4531B performs correctly.
The decimal point (Cf-01), minimum division (Cf-02) and rated
capacity (Cf-03) are set using the function mode.
The zero point input voltage (Cf-04), the span input voltage (Cf-05)
and the displayed value for the span input voltage (Cf-06) are
adjusted using the calibration mode.
Calibration setting using the function mode is also available. (Digital
span)
* During calibration, maintain a stable environment to prevent
calibration errors.
* When the measured value is stable, the HOLD LED is turned on.
* The decimal point blinks to indicate that the current value is not a
measured value.
7.1. Calibration Modes
In the measurement mode, press and to enter the
calibration mode.
Enter the zero point calibration mode.
Return to the measurement mode.
7.1.1. Zero point calibration mode
With nothing on the load cell, wait for
the HOLD LED to turn on and press .
Perform zero point calibration and
proceed to the span calibration mode.
Cancel zero point calibration and
proceed to the span calibration mode.
Hold down to display the mV/V value of
the zero point.
7.1.2. Span calibration mode
Input the value to be displayed when the actual load
for span calibration is applied to the load cell.
Wait for the HOLD LED to turn on and press .
Select the digit to be changed.
Increase the value of the digit to be changed.
Change the polarity.
Perform span calibration and
proceed to the storing mode.
Cancel span calibration and
proceed to the storing mode.
* After span calibration, the AD-4531B displays the mV/V value of
span calibration for 3 seconds, and then proceeds to the storing
mode.
7.1.3. Storing mode
Save the calibration zero, span and displayed
value acquired.
When calibration is not performed, data is not
saved.
Save the data acquired and return to the
measurement mode.
Do not save the data acquired and
return to the measurement mode.
7.2. Calibration Errors
Display Cause Remedy
C?e2
Voltage at zero point
calibration exceeds in the
positive direction.
C?e3
Voltage at zero point
calibration exceeds in the
negative direction.
Confirm the rating
and connection of the
load cell.
Display Cause Remedy
C?e4
The value of the calibration
weight exceeds the rated
capacity.
Use a proper calibration
weight.
C?e5
The value of the calibration
weight is less than the
minimum division value.
Use a proper calibration
weight.
C?e6
The load cell sensitivity is
insufficient.
Confirm the load cell
connection. Use a proper
calibration weight.
C?e7
Voltage at span calibration is less
than the voltage at the zero point.
Confirm the load cell
connection.
C?e8
The load cell output voltage is
too high when loaded to
capacity.
Use a load cell with a
greater rated capacity or
set a smaller rated
capacity value.
8. FUNCTION MODE
Use the function mode to set various functions. The set values are saved in
non-volatile memory and are maintained even if the power is disconnected.
8.1. Description of Functions
The first 2 digits of the function No. are the function group. The last 2 digits
of the Function No. are the function item.
Cf
Calibration function
f0
Basic function
f1
Comparator function
Use this function to set the comparator operation.
f2
Analog output function
Use this function to set the output values of the analog
voltage output and analog current output.
f3
,
f4
Serial communication function
Use this function to set the RS-232C and RS-485.
* Set the zero point input voltage (Cf-04), the span input voltage (Cf-05) and the
displayed value for the span input voltage (Cf-06) in the calibration mode.
* Set the upper limit value (f1-01) and lower limit value (f1-02) in the
comparator mode.
* When setting a function, the decimal point blinks to indicate that the
current value is not a measured value.
* In the digital filter setting mode (f0-02), press to confirm the
measured value.
When the measured value is displayed, the OK LED blinks.
Press to set the display to zero. And press to return to the
measurement mode.
8.2. Key Operation
In the measurement mode, press and to enter the function selection mode.
8.2.1. Function selection mode
Select the function group. (First 2 digits)
Select the function item. (Last 2 digits)
Enter the setting changing mode.
Save the setting in non-volatile memory and then return to the
measurement mode.
8.2.2. Setting changing mode (Two methods)
P Parameter selection method (All digits blinking)
Change the parameter.
Enter the setting and return to the function selection mode.
Cancel the setting and return to the function selection mode.
D Digital input method (Change the blinking digit only)
Move the digit to be changed to the right.
Change the value of the blinking digit.
Change the polarity.
Enter the setting and return to the function selection mode.
Cancel the setting and return to the function selection mode.
8.3. Function Items
8.3.1. Calibration (C function)
Function No.
Setting range
Function Description
Default value
Setting type
Cf-01
0 to 5
Decimal point
position
Decimal point position of the measured
value
0
:
000000 3
:
000.000
1
:
00000.0 4
:
00.0000
2
:
0000.00 5
:
0.00000
0
P
Cf-02
1 to 50
Minimum
division
Minimum division (d) of the measured value
1
:1
10
:10
2
:2
20
:20
5
:5
50
:50
1
P
Cf-03
1 to 999999
Rated capacit
Measurement is possible up to the value
of this setting plus 8 d (8 minimum
divisions)
Decimal point position depends on
Cf-01 .
70000
D
Cf-04
-7.00000
to 7.00000
Input voltage
of zero point
Input voltage from the load cell at zero
point
(Unit: mV/V)
0.00000
D
Cf-05
0.00001
to 9.99999
Input voltage
of span
Input voltage from the load cell at span
(measurement point - zero point)
(Unit: mV/V)
3.20000
D
Cf-06
-999999
to 999999
Displayed value
for input
oltage of span
Displayed value for span (measurement
point - zero point)
Decimal point position depends on
Cf-01 .
32000
D
Cf-07
0 to 100
Zero
adjustment
range
Range to enable zero adjustment by the
ZERO key
Expressed as a percentage of the rated
capacity with the calibration zero point
as the center.
100
D
Cf-08
0.0 to 5.0
Zero tracking
time
Performed in combination with zero
tracking width. (Unit: second)
When
0.0
, zero tracking is not performed.
0.0
D
Cf-09
0.0 to 9.9
Zero tracking
width
Performed in combination with zero
tracking time. (Unit: d)
When
0.0
, zero tracking is not performed.
0.0
D
Cf-10
0 to 2
Power on zero
Digital zero when the power is connected
0
: Digital zero function off
1
: Perform digital zero
2
: Use state when the power is disconnected
0
P
Cf-11
1 to 2
Zero operation
1
:
On with
,
Off with
+
2
: On or off with
+
1
P
8.3.2.
Basic Functions
Function No.
Setting range
Function Description
Default value
setting type
f0-01
00000
to 11111
Disable key
Each digit of the setting corresponds to a key
switch. Only available in the measurement mode.
Key assignment
0
:
Enabled
0 0 0 0 0
1
:
Disabled
00000
(Binary)
D
f0-02
0 to 9
Digital filter
Cutoff frequency
0
:
Off
5
: 2.8 Hz
1
: 11 Hz
6
: 2 Hz
2
: 8 Hz
7
: 1.4 Hz
3
: 5.6
8
: 1 Hz
4
: 4 Hz
9
: 0.7 Hz
8
P
f0-03
0 to 20
Display
update
rate
1
:
1 time/second
10
: 10 times/second
2
: 2 times/second
20
: 20 times/second
5
: 5 times/second
20
P
f0-04
0 to 4
Hold mode
0
:
Off
3
:
Bottom hold
1
:
Sample hold
4
:
Bipolar peak hold
2
:
Peak hold
1
P
f0-05
0.0 to 9.9
Hold
averaging
time
Set by the unit of 0.1 second.
When
0.0
, averaging is not performed.
0.0
D
f0-06
0000
to 1111
Latch
function
Corresponds to an external input latch.
Setting and latch
assignment
0
:
Off
0
0
0
0
1
:
On
Displayed value latch
Comparator latch
Analog output latch
Serial output latch
0000
(Binary)
D
f0-07
0 to 6
External
input 1
1
P
f0-08
0 to 6
External
input 2
0
:
Off
4
: Start hold
1
: DZ key
5
: Stop hold
2
: Hold
6
: Latch
3
: Print
2
P
f0-09
0 to 9
External
output 1
1
P
f0-10
0 to 9
External
output 2
0
:
Off
5
: OK
1
: DZ key
6
: Lo
2
: Holding
7
: Measuring (ON)
3
: Hold busy
8
: Measuring (1 Hz)
4
: Hi
9
: Measuring (50 Hz)
2
P
8.3.3. Comparator
Function No.
setting range
Function Description
Default value
setting type
f1-01
-999999 to
999999
Upper limit
value
Upper limit value of comparator.
Decimal point position depends on
Cf-01.
0
D
f1-02
-999999 to
999999
Lower limit
value
Lower limit value of comparator.
Decimal point position depends on
Cf-01.
0
D
f1-03
0 to 2
Comparato
mode
0
:Off
1
:On excluding the zero band
2
:Always on
2
P
f1-04
-999999 to
999999
Zero band
Set the zero band for the comparator
mode.
0
D
f1-05
1 to 3
Hysteresis
mode
Hysteresis direction
1
:Upward 2-level judgment
2
:Upper / lower limit judgment
3
:Downward 2-level judgment
2
P
f1-06
0.0 to 5.0
Hysteresis
time
Set the hysteresis time by the unit of 0.1 second.
When
0.0
, the hysteresis mode is not used.
0.0
D
f1-07
00 to 99
Hysteresis
width
Set the hysteresis width by the unit of d.
When
0.0
, the hysteresis mode is not used.
99
D
8.3.4. Analog
Function No.
setting range
Function Description
Default value
setting type
f2-01
-999999 to
999999
0 V
output
Measured value at DAV 0V output.
Decimal point position depends on
Cf-01.
0
D
f2-02
-999999 to
999999
10 V
output
Measured value at DAV 10V output.
Decimal point position depends on
Cf-01.
1000
D
f2-03
-999999 to
999999
4 mA
output
Measured value at DAI 4 mA output.
Decimal point position depends on
Cf-01.
0
D
f2-04
-999999 to
999999
20 mA
output
Measured value at DAI 20 mA output.
Decimal point position depends on
Cf-01.
1000
D
8.3.5. Serial Communication
Function No.
Setting range
Function Description
Default value
Setting type
f3-01
2400 to
38400
Baud rate
2400
: 2400 bps
19200
: 19200 bps
4800
: 4800 bps
38400
: 38400 bps
9600
: 9600 bps
2400
P
f3-02
7 to 8
Data bit
length
7
:7 bits
8
:8 bits
7
P
f3-03
0 to 2
Parity
0
:None
1
:
Odd
2
:
Even
2
P
f3-04
1 to 2
Stop bit
1
:1 bit
2
:2 bits
1
P
f3-05
1 to 2
Terminator
1
:CRLF
2
:CR
1
P
f3-06
1 to 6
Output
mode
1
:
Stream mode
2
:
Manual print mode
3
:
Auto print mode (Outputs data once
when the measured value exceeds
the zero range and is stabilized for
the first time.)
4
:
Auto print mode (Outputs data each
time the measured value exceeds
the zero range and is stabilized.)
5
:
Command mode
6
:
Jet stream mode (Outputs data at
each sampling, depending on the
baud rate.)
2
P
f3-07
00 to 99
Instrument
No.
ID that is added to the serial output
When
0.0
, the ID is not added.
00
D
8.3.6. Unit
Function No
Setting
range
Function Description
Default value
Setting type
f4-00
0 to 4
Unit
0
: Specify the unit character
(f4-01 to f4-05)
1
: kg
3
: t
2
:
g
3
:Ib
1
P
f4-01
f4-02
f4-03
f4-04
f4-05
00 to 7F
Unit character 1
Unit character 2
Unit character 3
Unit character 4
Unit character 5
Unit character added to the serial output.
Set using the hexadecimal ASCII code.
00
: none
00
(
Hexadecimal
)
D
Measurement
mode
Measuring
Measuring
Example of
the specified value