
45
Note
The upper/lower limit values can be confirmed by external commands using the RS-232C serial
interface.
e.g., Command ?HH (to confirm secondary upper limit value)
Response HH, + 100.00000 g (secondary upper limit value=100.00000 g)
Command ?HI (to confirm upper limit value)
Response HI +080.00000 g (upper limit value=80.00000 g)
Command ?LO (to confirm lower limit value)
Response LO, +060.00000 g (lower limit value=60.00000 g)
Command ?LL (to confirm secondary lower limit value)
Response LL, +040.00000 g (secondary lower limit value=40.00000 g)
indicates a space (20h)
Adding the comparison results (AD-4212A only)
To add the comparison results to the output data, set “Comparison results (Cp-r)” parameter of the
function table to “1”.
ST, H H , 110 .0000 g HH (Only when OP-04 is installed.)
ST, HI, +090.0000 g HI
ST, OK,+070.0000 g OK
ST, LO, +050.0000 g LO
ST, LL,+030.0000 g LL (Only when OP-04 is installed.)
ST, - -,+000.0000 g No comparison when unstable or near zero
indicates a space (20h)
Notes
Select A&D standard format. Set “Data format (type)” parameter of the function table to “0”.
The comparison results can not be added to the weighing data stored in memory.
Kommentare zu diesen Handbüchern