A&D EK-09H Spezifikationen

Stöbern Sie online oder laden Sie Spezifikationen nach Messen, Testen & Steuern A&D EK-09H herunter. A&D EK-09H Specifications Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1

x 1WMPD4000465B

Seite 2 - COMPLIANCE WITH FCC RULES

85-4. Selecting a weighing unit Press the MODE key to select a unit. The following sections are a description of the three common units: g (g

Seite 3 - CONTENTS

95-6. Weighing range for the EW-i series  The EW-i series have three weighing ranges, and the display shows which range the weight value belongs to

Seite 4

105-7. Counting mode (pcs) Determines the number of objects in a sample. Calculates the reading, using the basic sample unit weight, and determines

Seite 5 - 2. UNPACKING

11 8. After setting the unit weight at step 6, add a few more samples on the pan. Then, the ACAI mark will turn on. (Add more than three pieces. The

Seite 6 - 3. PART NAMES AND FUNCTIONS

126. COMPARATOR The results of the comparison are indicated by HI, OK or LO on the display. The comparison is as follows: LO < Lower limit value

Seite 7 - 4. SETTING UP

13Entering the upper and lower limit values 6. With Cp Hi displayed, press the PRINT key. Enter the upper limit value using the following key

Seite 8 - 5. OPERATION

147. CALIBRATION This function adjusts the balance for accurate weighing. Perform calibration in the following cases.  When the balance is first

Seite 9 - 5-3. Units

15 5. At step 3, pressing the PRINT key weighs the zero-point value. Do not touch the pan during weighing. When the zero calibration is completed,

Seite 10 - 150.3

167-2. Gravity acceleration correction When the balance is first used or has been moved to a different place, it should be calibrated using a cali

Seite 11 - Operation

178. FUNCTIONS 8-1. Key operation 8-2. Entering the function setting mode In the weighing mode, press and hold the SAM

Seite 12 - 5-7. Counting mode (pcs)

© 2011 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into an

Seite 13 - 100 0

188-3. Setting example To set auto power-off function to “Enabled”, and the ACAI function to “Disabled”. 1. Press and hold the SAMPLE key to

Seite 14 - 6. COMPARATOR

198-4. Storing weighing units It is possible to store the weighing units that will be actually used from the units available. For the units availabl

Seite 15

208-5. Function list Class Item Param- eter Description  0 Auto power-off disabled poff Auto power-off 1 Auto power-off enabled Automatically powe

Seite 16 - 01200.0

21 Class Item Param- eter Description  0 7 bits, even parity 1 7 bits, odd parity btpr Data and parity 2 8 bits, non parity 0 ACAI disabled a

Seite 17 - 1000.0

229. RS-232C SERIAL INTERFACE This interface allows the EK/EW-i series to be connected with a multifunction printer or a personal computer.  The

Seite 18 - 9.7980

239-2. Data format S T , + 0 0000.00 g CR LF Header Data Unit Terminator  There are four types of headers: ST : Stabl

Seite 19 - 8. FUNCTIONS

24Key mode Set the function “prt 1, 2 or 3”. When the PRINT key is pressed while the weight data is stable (the stability mark is on), the bal

Seite 20 - 8-3. Setting example

2510. ID NUMBER AND GLP The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used. The following GLP data is transmi

Seite 21 - 8-4. Storing weighing units

2610-2. GLP report  To print the GLP report to the AD-8121B, select the function setting “info 1” and “pUse 1” for the balance, and use MODE 3 of th

Seite 22 - 8-5. Function list

27Calibration test report The calibration test mode is used to compare a calibration test weight with the calibration test data weighed by the balanc

Seite 23

1CONTENTS 1. INTRODUCTION ...3 2. UNPACKING

Seite 24 - 9. RS-232C SERIAL INTERFACE

28 AD-8121 format “info 1” General format “info 2” : Space, ASCII 20h CR: Carriage return, ASCII 0Dh LF: Line feed, ASCII 0D

Seite 25 - 9-3. Data output mode

29End block 3. Press and hold the PRINT key. Release the PRINT key when recend is displayed. The balance outputs the “End block”. 4.

Seite 26 - 9-4. Command mode

3011. OPTIONS The following options are available for the EK/EW-i series:  OP-04 Comparator relay output and buzzer  OP-07 Underhook assembly

Seite 27

31 OP-04 Installation 1. Remove the cover of the option slot on the rear of balance by pressing and lowering it down. 2. Insert the option into the

Seite 28 - 10-2. GLP report

3211-3. OP-09 Rechargeable battery pack (Ni-MH) By installing the rechargeable battery pack (Ni-MH) into the balance, cordless operation can be carr

Seite 29

3312. MAINTENANCE 12-1. Notes on maintenance  Do not disassemble the balance. Contact your local A&D dealer if your balance needs service or r

Seite 30 - 123.4

34 CAL errors Warning to indicate that calibration has been canceled because the calibration weight is too heavy. Warning to indicate that calibrati

Seite 31 - recend

3513. SPECIFICATIONS 13-1. EK-i series MODEL EK-120i EK-200i EK-300i EK-600i Weight capacity 120 g 200 g 300 g 600 g Min. display “d” 0.01 g

Seite 32 - 11. OPTIONS

36 MODEL EK-410i EK-610i EK-4100i EK-6100i Weight capacity 400 g 600 g 4000 g 6000 g Min. display “d” 0.01 g 0.01 g 0.1 g 0.1 g No. of sample

Seite 33

3713-3. Other weighing units MODEL EK-120i EK-200i EK-300i EK-600iEK-1200i EK-2000i EK-3000i EK-6000i EK-12KiCapacity 4.2330 7.0550 10.5820 21.165 4

Seite 34 - 11-4. OP-12 Carrying case

211. OPTIONS...30 11-1. OP-04 Compa

Seite 35 - 12-2. Error codes

3813-4. Dimensions EK-120i / EK-200i / EK-300i EK-600i / EK-1200i / EK-2000i EK-3000i / EK-6000i / EK-12Ki EK-410i / EK-610i

Seite 36

39GRAVITY ACCELERATION MAP Values of gravity at various locations Amsterdam 9.813 m/s2 Manila 9.784 m/s2 Athens 9.800 m/s2 Melbourne 9.800

Seite 38 - 13-2. EW-i series

3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013 JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148 A&D ENGINEERING, I

Seite 39 - 13-3. Other weighing units

31. INTRODUCTION This manual describes how this balance works and how to get the most out of it in terms of performance. The EK-i and EW-i series b

Seite 40 - 13-4. Dimensions

43. PART NAMES AND FUNCTIONS Turns the power on or off. Held down to enter the functionsetting mode. Weighing

Seite 41 - GRAVITY ACCELERATION MAP

54. SETTING UP 4-1. Setting up your balance 1. Place the weighing pan on the main unit as shown on the previous page. 2. Adjust the level of the ba

Seite 42 - World map

65. OPERATION 5-1. Turning the power ON and OFF 1. Press the ON/OFF key to turn the power ON. All of the symbols are displayed as sh

Seite 43

75-3. Units The most common unit of weight used around the world is the gram, but there is often a need to shift to alternative units specific to th

Kommentare zu diesen Handbüchern

Keine Kommentare